The actual sample was not returned for evaluation.Spinefrontier investigation into this matter included design of an experiment using components from the same lot used to manufacture the complainant device.The failure mode inserter becomes disengaged during impaction was successfully re-created based on the following observation (knob loosening and eventually disengaging from the shaft component).Signs of wear was evident on both the shaft and knob components.However, the failure mode only replicated in scenarios when a worn knob was used with a new shaft, and in scenarios when over-tightening of the inserter onto the implant.The results of the experiment concluded that the failure was isolated to only the knob component and only occurred in scenarios where overtightening was a factor.A review of the current plift fmea revealed disassembly is known risks associated with the inserter with a rpn value of 6, which is less than actionable risk level.A review of the p-lift inserter labeling establishes there are adequate labeling with regards to over-tightening.Component wear is mitigated through pre/post-surgery inspection.
|