ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100R
Device Problem Premature Discharge of Battery (1057)
Patient Problems Headache (1880); Failure of Implant (1924)
Event Date 03/13/2019
Event Type  malfunction  
Event Description
The recipient reportedly elected explant surgery due to poor battery life.The recipient presented with headaches.The recipient's device was explanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the fantail was damaged near the epoxy header, and the electrode was sliced along the lead prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection confirmed broken electrode wires at the epoxy header.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.Voltage values at some nodes were elevated.Internal inspection did not reveal any anomalies on internal components.This older device configuration is not currently manufactured.This is the final report.
 
Manufacturer Narrative
The recipient was reimplanted with another cochlear device.Additional information regarding recipient status will not be provided.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key8624446
MDR Text Key145626098
Report Number3006556115-2019-00277
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/15/2001
Device Model NumberAB-5100R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/20/2019
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received06/24/2019
07/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age28 YR
-
-