The reported event is inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿adhesive /hydrocolloid contaminated/dirty¿.A potential root cause for this failure mode could be ¿material handling¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity." the device was not returned.
|