Model Number 9770501 |
Device Problems
Battery Problem (2885); Charging Problem (2892)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per tm, battery will not hold a charge.'we currently have to keep the same us plugged in or it will turn off at 68% battery life.'.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the battery will not hold a charge is confirmed.The root cause is an internal failure in the lithium-ion battery.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per tm, battery will not hold a charge.'we currently have to keep the same us plugged in or it will turn off at 68% battery life.'.
|
|
Event Description
|
Per tm, battery will not hold a charge.'we currently have to keep the same us plugged in or it will turn off at 68% battery life.'.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the battery will not hold a charge is confirmed.The root cause is an internal failure in the lithium-ion battery.The device was serviced, tested and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|
|