• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3361-40C
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332)
Patient Problems Syncope (1610); Bradycardia (1751); Dizziness (2194)
Event Date 05/08/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
 
Event Description
During an in-clinic visit due to dizziness and bradycardia, inductive telemetry issues were observed and the device was unable to communicate with the merlin programmer.The physician suspected premature battery depletion, and the device was successfully replaced.The patient was stable following the procedure with no additional adverse consequences.
 
Manufacturer Narrative
Additional information: as received, the device did not communicate.The cause of the no communication was found to be due to low battery voltage from premature depletion.Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ABBOTT
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8638204
MDR Text Key145994442
Report Number2938836-2019-03946
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Model NumberCD3361-40C
Device Lot Number4710513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/08/2019
Initial Date FDA Received05/23/2019
Supplement Dates Manufacturer Received01/28/2020
Supplement Dates FDA Received01/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-