During an in-clinic visit due to dizziness and bradycardia, inductive telemetry issues were observed and the device was unable to communicate with the merlin programmer.The physician suspected premature battery depletion, and the device was successfully replaced.The patient was stable following the procedure with no additional adverse consequences.
|
Additional information: as received, the device did not communicate.The cause of the no communication was found to be due to low battery voltage from premature depletion.Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|