The device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The balloon could not be inflated due to a leak in the shaft polymer extrusion of the device.Visual and microscopic examination of the balloon material identified no issues which could potentially have contributed to this complaint.Visual and microscopic examination observed no damage to the blades.All blades were present and fully bonded to the balloon surface.The markerbands and tip section of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.The returned device was attached to a boston scientific encore inflation unit.Positive pressure was applied, and liquid was observed to be leaking from a shaft longitudinal tear beginning approximately 13mm distal of the guidewire port and extending approximately 2mm distally along the shaft.No other issues were identified during the product analysis.Visual and tactile examination found that the hypotube was kinked at several locations along its length.No other issues were identified during the product analysis.
|
Reportable based on device analysis completed on 03may2019.It was reported that difficulty inflation of the balloon occurred.A target lesion was located at mid right coronary artery.A 15mmx3.50mm wolverine coronary cutting balloon monorail.During procedure, at inflation, it was noted that the device lost pressure at 2, 4, then 6 atmospheres.Subsequently, they revealed that there was a leak at the shaft transition not at the balloon.The procedure was completed with a non bsc device.No complications reported.However, device analysis revealed aft longitudinal shaft tear.
|