On (b)(6) 2019, a 19mm regent valve was selected for implant.While implanting the device, the physician observed a leaflet that was not coapting properly and appeared to be stuck.The physician explanted the valve and implanted a 17mm regent valve in the smaller annulus.No patient consequences were reported.
|
On (b)(6) 2019, a 19mm regent valve was selected for implant.While implanting the device, the physician observed a leaflet that was not coapting properly and appeared to be stuck.The physician explanted the valve and implanted a 17mm regent valve in the smaller annulus.No patient consequences were reported.
|
The reported event of an obstructed leaflet could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined; however, information from the field indicated that a smaller, 17mm, regent valve was used to complete the procedure, which is possible evidence of initial over-sizing.Please note, per the instructions for use artmt100122074, rev a, "proper valve size selection is crucial.Do not oversize the valve.If the native annulus measurement falls between two sjm regent¿ mechanical heart valve sizes, use the smaller size sjm regent¿ mechanical heart valve.".
|