Model Number ROTAFLOW DRIVE
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/07/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).
 
Event Description
It was reported that the rfd displays head error when rpm is increased above 1500 and is non-functional.Issue was observed at the mahwah depot and no issues were reported from the field.It is unknown how, when or where this failure occurred.The described failure can lead to a pump stop.No patient involvement.Complaint onetrack; (b)(4).
 
Manufacturer Narrative
The reported failure head error was investigated in germany.The rotaflow drive has been sent to the manufacturer emtec for investigation and repair, the outcome of the investigation was, that the could reproduce the failure and repaired the drive.Thus the reported failure "head error" could be confirmed.Most possible root cause could be also determined as: 1.The head error is caused by the hot plug.When device is in operation and the power plug is plugged in or out.And this leads to a damage at the control board of the rotaflow.2.The head error follows by the sig error.This is when the ultrasonic crème is applied to the flow bubble sensor.Then the centrifugal pump is causing backflow and this leads to the head error.3.The head error is also caused by shaking the drive.This is when the motor (which is controlled by the optical tacho) is not blocked when adjusting to 0 then it could lead to error when slight shaking.The motor could then slightly rotate.4.The head error can be caused by connection issues between the console (rfc) and the drive (rfd).This is when the cable connection is disturbed by defective pins.It is unknown how, when or where this failure occurred.Since the fault and possible root causes are already known, there is a relationship between the device and the complaint.The occurence rate is below the acceptance rate, thus no remedial action required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint onetrack#(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8645373
MDR Text Key146240571
Report Number8010762-2019-00143
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW DRIVE
Device Catalogue Number701022161
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-