Device evaluation by manufacturer: a field service engineer (fse) arrived at the user facility and confirmed the reported event by observing that the sampling nozzle was completely clogged.The fse proceeded to perform sampling nozzle flush with bleach, followed by di water, alcohol, and air.The fse then adjusted the pressure board per manufacturer's specifications.The aia-2000 was validated by running quality control on fsh, which passed within published ranges.No further action was required.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 07-apr-2018 through aware date 07-may-2019.There were no other similar events reported during the searched period.The aia-2000 operator's manual under a4.Appendix 4: error messages, states the following: 2070 clogging detected during specimen suction by main arm.Cause: the negative pressure detected after specimen suction exceeded the standard.The specified amount of specimen may not have been obtained because the sampling nozzle was blocked.The measurement result will be flagged (sc flag).Solution: verify that the specimen is free of solid substances (such as fibrin) or that there is sufficient volume of specimen if it is prepared in the primary tube, and retry the measurement.If retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event was due to a clogged sampling nozzle.
|
A customer reported getting error message "2070 clogging detected during specimen suction by main arm" while running patient samples on the aia-2000 instrument.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient samples for intact parathyroid hormone (ipth), follicle stimulating hormone (fsh), beta human chorionic gonadotropin (bhcg), and prolactin (prl).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|