Model Number TLEAD1058-70B |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 04/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis of the device is currently in progress.The manufacturing records were reviewed and no issues were found.
|
|
Event Description
|
It was reported to nevro that during the trial procedure, the physician encountered resistance while placing a lead, possibly due to scar tissue, resulting in two electrodes becoming detached from the lead.There were no injuries to the patient and the physician decided to leave the detached electrodes in place.The patient completed the trial with effective pain relief and the physician plans to remove the detached electrodes during the permanent implant procedure.
|
|
Manufacturer Narrative
|
Follow-up indicated the patient had the permanent implant procedure and the detached electrodes were removed.The patient is currently using their implanted device with complete pain relief.The lead was returned for analysis.Two metal electrodes were missing on the distal end of the lead and a conductor wire was protruding.A knick was found at the lead tip and an indentation was observed on the lead body.Sem analysis showed each of the frayed wire strands were consistently necked and fractured.The indentation on the lead body and knick found on the tip indicate it was likely caught on a sharp edge.The morphology indicates that this was likely caused by a sharpened tip of an insertion needle.A needle likely caught the surface of the lead during lead withdrawal, leading to a cut along the surface up to the missing electrodes.A significant amount of force was then applied, ultimately leading to the detachment of the metal part of the two contacts.
|
|
Search Alerts/Recalls
|