B.BRAUN SURGICAL SA PREMILENE 4/0 (1.5) 90CM 2XHR17 CV; CARDIAC SUTURE Back to Search Results
Model Number C2090902
Device Problems Material Separation (1562); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).When additional information is received a follow up report will be submitted.
 
Event Description
It was reported the thread and needle detached.The reporter indicated that the needle detaches from the thread as soon as you try to use it.There is no patient data available.No other information has been provided.
 
Manufacturer Narrative
Investigation: samples received: none.Analysis and results: there are no previous complaints of the same code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample for analysis.Without any closed sample we cannot carry out an analysis in order to take a decision.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyze it.Please note that when no samples are received our analysis is very limited.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.No corrective/preventive actions needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMILENE 4/0 (1.5) 90CM 2XHR17 CV
Type of Device
CARDIAC SUTURE
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key8655226
MDR Text Key146535985
Report Number3003639970-2019-00438
Device Sequence Number1
Product Code GAW
Combination Product (y/n)N
PMA/PMN Number
K980703
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC2090902
Device Catalogue NumberC2090902
Device Lot Number116431
Was Device Available for Evaluation? No
Distributor Facility Aware Date05/22/2019
Date Manufacturer Received07/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-