BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Catalog Number 103401 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The event occurred on an unspecified date in (b)(6) 2019.Address: (b)(6).The actual device was not available; however, one photograph of the sample was provided for evaluation.The visual inspection of the provided picture showed a leak in the red access site (arterial injection port).The reported condition was verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The cause of the condition was not determined.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that as soon as treatment started with a gambro cartridge set, an external blood leakage was observed at the arterial injection port.Treatment was discontinued and the blood was returned to the patient.The blood loss was reported as ¿a very small amount of blood leaked¿.The set was replaced and treatment was restarted and completed without any issues.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|