Model Number SCD396 |
Device Problems
Material Fragmentation (1261); Detachment of Device or Device Component (2907); Intermittent Energy Output (4025)
|
Patient Problems
No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
|
Event Date 05/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the prostatectomy, when all the pieces of the scissors were mounted, the transducer showed a green light indicating that it was fit to use.After approximately 30 minutes of surgery, the dissector presented the red light at some times that was triggered.All the pieces were removed and replaced as per manufacturer's instructions, however the transducer continued to show red light more and more frequently until the scissors tip broke and fell into the cavity.The scissor was replaced by another dissector.
|
|
Event Description
|
According to the reporter, during the prostatectomy, when all the pieces of the scissors were mounted, the transducer showed a green light indicating that it was fit to use.After approximately 30 minutes of surgery, the dissector presented the red light in some times that was triggered.All the pieces were removed and replaced as per manufacturer's instructions, however the transducer continued to show red light more and more frequently until the scissors tip broke and fell into the cavity.It was recovered, but not forwarded due to contamination level.The scissor was replaced by another dissector.No patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Evaluation summary: one sonicision cordless ultrasonic dissector was returned for evaluation.The returned product did not meet specification as received.Visual inspection of the disposable device revealed that the tip of the waveguide was broken off.The broken piece was not returned.Investigation personnel performed functional testing on the returned hand piece using a test lab generator and battery.The assembled device returned a green light and a welcome tone, but immediately returned a red led indicator with an error tone (alarm activation) when the device was activated.This characteristic indicated that the device was not functional.The waveguide had its tip missing.Investigation personnel concluded that the titanium waveguide was in use when it cracked and broke off, and may have come in contact with any of the following; hemostats, clips, staples, retractors, etc.During use.The device instructions for use (ifu) currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.Contact with metal objects during use will cause the active blade to crack and may eventually break off.This issue is specific to ultrasonic dissectors.The ifu advises device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the prostatectomy, when all the pieces of the scissors were mounted, the transducer showed a green light indicating that it was fit to use.After approximately 30 minutes of surgery, the dissector presented the red light in some times that was triggered.All the pieces were removed and replaced as per manufacturer's instructions, however the transducer continued to show red light more and more frequently until the scissors tip broke and fell into the cavity.It was recovered, but not forwarded due to contamination level.The scissor was replaced by another dissector.
|
|