AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550350-12
Device Problems Difficult to Remove (1528); Material Too Rigid or Stiff (1544); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported the procedure was performed to treat a lesion with no calcification and moderate tortuosity in the right coronary artery.The lesion was pre-dilated with an unknown balloon catheter.Resistance was noted during advancement of the 3.5 x 12 mm xience sierra stent delivery system (sds) with the guiding catheter.The sds was able to reach the target lesion and the stent was deployed successfully.During removal of the sds, resistance was felt again with the guiding catheter; however, the sds was able to be removed independently.The physician commented that the shaft of the sds felt stiffer than normal.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).Evaluation summary: the device was returned for analysis.The reported difficulty to position and difficulty to remove could not be confirmed due to the condition the returned device.The reported device operates differently than expected could not be confirmed as it is an opinion of the physician.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the device interacted with the guiding catheter during advancement and removal causing the reported difficulty to position and difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8658881
MDR Text Key146668955
Report Number2024168-2019-04259
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2020
Device Catalogue Number1550350-12
Device Lot Number9011741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2019
Initial Date Manufacturer Received 05/09/2019
Initial Date FDA Received05/31/2019
Supplement Dates Manufacturer Received06/20/2019
Supplement Dates FDA Received06/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-