Model Number MODEL 100
Device Problem Display Difficult to Read (1181)
Patient Problem No Patient Involvement (2645)
Event Date 05/06/2019
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint of 'autopulse platform system (sn (b)(4)) liquid crystal display (lcd) screen did not display any data' was confirmed during functional test.The most probable root cause is due to damage processor board.During visual inspection, there was not visual damage observed on the returned autopulse platform.Functional test was unable to be performed, the autopulse platform displayed a blank screen upon powering up the device.The processor board was replaced to correct the fault.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
It was reported that when the autopulse platform system is powered up the liquid crystal display (lcd) screen did not display any data or text.The user tried multiple times with no results.This was noted in-house, no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key8660587
MDR Text Key147238283
Report Number3010617000-2019-00440
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-