The service protocol has not been received despite several requests.This complaint will be closed without the service protocol.Therefore it cannot be assured in which technical status this device is currently.This complaint will be reopened and appropriate actions will then be taken if the service protocol is received and contains any further relevant information.And therefore no most probable root cause possible.For this error an nc was opened and investigated: in the course of the investigation of nc-17-06-011 all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system see rotaflow system user manual, mcv-ga-10000703-de-11, contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noticeable that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|