ZIMMER BIOMET, INC. DRILL PIN AND SCREW INSERTER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a total knee procedure, while trying to insert a pin, the instrument fractured.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned screw inserter/extractor confirms that a small portion near the distal end of the device has fractured off.It was also noted to exhibit signs of repeated use (surface scratches).Device history record was reviewed and no discrepancies were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL PIN AND SCREW INSERTER
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8664677
MDR Text Key147052090
Report Number0001822565-2019-02303
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00590102100
Device Lot Number63232981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2019
Initial Date FDA Received06/03/2019
Supplement Dates Manufacturer Received10/02/2019
Supplement Dates FDA Received10/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-