MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550300-15
Device Problems Tear, Rip or Hole in Device Packaging (2385); Material Twisted/Bent (2981)
Patient Problem No Patient Involvement (2645)
Event Date 05/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that while unpacking the device, the grey pouch was twisted and connected with the clear sterile pouch.They could not be separated and device sterility was not known.The device was not used and there was no patient involvement.No additional information was provided regarding this device issue.
 
Manufacturer Narrative
Internal file number - (b)(4).If follow-up what type, correction: mfr site-reg#.Evaluation summary: the device was returned for analysis.The reported packing issues and packaging damage were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported packaging issues and packaging damage appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8664739
MDR Text Key146979593
Report Number2024168-2019-04344
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/21/2020
Device Catalogue Number1550300-15
Device Lot Number9021541
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/03/2019
Initial Date Manufacturer Received 05/15/2019
Initial Date FDA Received06/03/2019
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received08/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-