(b)(4).Received one set of connector and catheter connected to filter.The clip was attached to connector, fixed by medical tape, at slightly shifted position.Visual inspection: catheter was crushed at a position of 922mm from tip.Connector was identical to standard sample, and no abnormality was found on latch portion.Clip was identical to standard sample, although the slit is wider, which is considered to be due to usage.Dimension check: the length of catheter was 1017mm, which satisfies spec of 1002.5-1017.5mm.The outer diameter of catheter tip was 0.8551mm, satisfies spec of 0.80-0.88mm.The slit width of the clip was 1.3287mm, satisfies spec of 0.5-1.5mm.Functional test: tensile test with returned connector and unused catheter was conducted.(returned catheter could not be used since disinfection could not be performed due to crush) the measured tensile strength was 9.4n which satisfies spec of >5.5n.When catheter was inserted to the connector, catheter was smoothly inserted up to the target position, and the cover of connector was locked securely.The insertion depth is confirmed to satisfy specified standard.Clip could not be attached securely to the connector for the first three times, but after fourth trial, it could be secured.Production acceptance inspection history: acceptance inspection record of catheter, connector, and clip of the subject batches was examined, and no abnormalities were found.The product design is being updated to increase the force required to open the catheter connector under change control: hc-chc-m-div-1306.This report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|