Model Number SCD396 |
Device Problems
Material Fragmentation (1261); Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/07/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned, however, a picture was provided by the customer for analysis.The photo of the disposable device revealed that the tip of the waveguide was broken off.The broken piece was shown in the picture.The reported condition was confirmed, based on the picture received.According to the picture, the waveguide had its tip broken.Investigation personnel concluded that the titanium waveguide was in use when it cracked and broke off, and may have come in contact with any of the following; hemostats, clips, staples, retractors, etc.During use.The investigation identified the cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.Contact with metal objects during use will cause the active blade to crack and may eventually break off.This issue is specific to ultrasonic dissectors.The instructions for use (ifu) advises device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Ifu also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure in endometriosis, at the end of the surgery the light turned red and when they removed the device from patients cavity, the jaw was broken.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a procedure in endemetriosis, at the end of the surgery the light turned red and when they removed the device from patients cavity, the jaw was broken.There was nothing fell into patients cavity.They completed the procedure using a permanent bipolar forceps.There was no patient injury.
|
|