Model Number 2426-0007 |
Device Problem
Reflux within Device (1522)
|
Patient Problem
Missed Dose (2561)
|
Event Date 05/06/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that the nurse observed the secondary mini-bag medication back flow into the primary line while priming the tubing set.Although the issue occurred during priming, this resulted in the patient not receiving the ordered medication gravol.The customer further stated that there was no adverse effects cause to the nurse or the patient.
|
|
Event Description
|
It was reported that the nurse observed the secondary mini-bag medication back flow into the primary line while priming the tubing set.Although the issue occurred during priming, this resulted in the patient not receiving the ordered medication gravol.The customer further stated that there was no adverse effects cause to the nurse or the patient.
|
|
Manufacturer Narrative
|
The customer¿s report of secondary set back flowed into primary was not confirmed.Visual inspection of the set noted no damage or any anomalies.The check valve was visually inspected under magnification.The check valve was noted to be assembled in the set correctly, and the silicone membrane was centered.Functional testing resulted in no air bubbles or fluid going past the check valve or into the primary bag.Disassembly of the check valve resulted in normal findings.No particulates were noted on the diaphragm membrane or any solvent to the check valve component with no damages to the interior of the check valve or to the diaphragm.The root cause of this failure was not identified.
|
|
Search Alerts/Recalls
|