Model Number DM3500
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the implantable cardiac monitor was found not working by supplier before it was supplied to hospital.Device was returned for analysis.No patient was involved.
 
Manufacturer Narrative
Device was returned due error message alert being triggered and would not allow further interrogation.Analysis of device image indicated that memory corruption occurred during the manufacturing process.Further testing was performed by reloading the product code and all tests results were normal, memory corruption could not be reproduced.A manufacturing anomaly may have occurred that resulted in the reported error message due to memory corruption.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8675958
MDR Text Key147239671
Report Number2017865-2019-09175
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/24/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000056777
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2019
Was the Report Sent to FDA? No
Date Manufacturer Received09/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-