• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC Back to Search Results
Catalog Number 8065751657
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A surgeon reported that the valved trocars leaked during a procedure.There was no patient harm.
 
Manufacturer Narrative
Three opened 23 gauge trocar cannula/hub assemblies were received for evaluation.The returned samples were visually inspected and were found to be non-conforming with damage to the septum.Leak testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates there are four additional complaints associated with the lot for the reported issue.The exact root cause for this complaint is unknown, however, a potential contributing factor related to the manufacturing process of the valves has been identified.An investigation has been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS VALVED ENTRY SYSTEM
Type of Device
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8676490
MDR Text Key147578352
Report Number2028159-2019-01037
Device Sequence Number1
Product Code NGY
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Catalogue Number8065751657
Device Lot Number17030971X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2019
Initial Date Manufacturer Received 05/22/2019
Initial Date FDA Received06/06/2019
Supplement Dates Manufacturer Received08/20/2019
Supplement Dates FDA Received08/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-