ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC
|
Back to Search Results |
|
Catalog Number 8065751657 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A surgeon reported that the valved trocars leaked during a procedure.There was no patient harm.
|
|
Manufacturer Narrative
|
Three opened 23 gauge trocar cannula/hub assemblies were received for evaluation.The returned samples were visually inspected and were found to be non-conforming with damage to the septum.Leak testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A complaint history examination indicates there are four additional complaints associated with the lot for the reported issue.The exact root cause for this complaint is unknown, however, a potential contributing factor related to the manufacturing process of the valves has been identified.An investigation has been completed and manufacturing process enhancements have been implemented in order to improve the performance of the valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
|