• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790030
Device Problems Fracture (1260); Loss of Osseointegration (2408)
Patient Problems Inadequate Osseointegration (2646); No Code Available (3191)
Event Date 05/20/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised to address loosening of the metaglene component at bone to implant interface.One of the metaglene screws was broken.It had been revised once before, from the original surgery on (b)(6) 2016.Surgeon removed the glenosphere, metaglene and screws.He left the broken part of the screw in the glenoid.He re-reamed the glenoid, changed the positioning of the metaglene, went up to a 42 glenosphere, used an eccentric head, put 4 new screws in, and used a +9 spacer and a 42 +3 cup.Doi: (b)(6) 2016, dor: (b)(6) 2019, left shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. h10 additional narrative:   added: h6 (device) corrected: d1, d4, h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X30MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key8679783
MDR Text Key147360048
Report Number1818910-2019-95477
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027850
UDI-Public10603295027850
Combination Product (y/n)N
PMA/PMN Number
K120174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130790030
Device Lot Number5265634
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/20/2019
Initial Date FDA Received06/07/2019
Supplement Dates Manufacturer Received07/09/2019
07/16/2019
Supplement Dates FDA Received07/10/2019
07/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-