U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number VJ4304TS_
Device Problems Device Damaged Prior to Use (2284); Contamination /Decontamination Problem (2895); Device Contamination with Chemical or Other Material (2944)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 05/15/2019
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.
 
Event Description
It was reported that an unknown black particulate was observed inside of the connector in an opened pouch before use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
The ir spectrum of the black particulate noted in the disposable pressure transducer male luer connector could not be determined due to its microscopic size coupled with low absorbance energy.
 
Manufacturer Narrative
One single dpt-vamp jr.Kit with pressure tubing was returned for examination.The reported event of ¿unknown black particulate was observed inside the connector¿ was confirmed.Several unknown black materials were observed on the male luer of the dpt zero-stopcock located inside of the connection to the female connector of the pressure line.The largest material was approximately 1.0 x 0.5 mm in size, and the others were less than 0.5 mm in size.The materials stayed at the same location on the male luer after 5 minutes of continuous flushing.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.A supplemental report will be forthcoming when the chemistry results are completed.It is common clinical practice to inspect all products before usage.Additionally, these products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.In this case, the particulate was noticed before use.Chemistry results are pending.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
The investigation of the unknown black particulate found inside the connector of a dpt (disposable pressure transducer) found that the potential root cause for this nonconformance could be a supplier material defect.As this defect cannot be generated at our manufacturing process, the supplier was notified.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key8680584
MDR Text Key147878735
Report Number2015691-2019-02074
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/07/2023
Device Model NumberVJ4304TS_
Device Lot NumberKP0680MT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2019
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-