Model Number 459878 |
Device Problems
Electrical /Electronic Property Problem (1198); High impedance (1291); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Muscle Stimulation (1412)
|
Event Date 05/05/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there an alert for out of range, high left ventricular (lv) lead impedance, which led to a polarity switch to a unipolar configuration.It was noted the patient experienced extracardiac stimulation in the unipolar configuration.The lv lead was reprogrammed and remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information reported the lv lead was programmed off.
|
|