ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX XL LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1158T/86
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented for normal generator changeout procedure.During post procedure interrogation, the left ventricular lead exhibited noise.Noise was reproducible with a stress test.It was suspected that the lead had an insulation breach and externalized conductors.Chest x-ray and fluoroscopy were performed but insulation breach could not be identified.Programming changes were made.The patient would continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX XL LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8684264
MDR Text Key147500930
Report Number2017865-2019-09389
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2014
Device Model Number1158T/86
Device Lot Number3478491
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-