It was reported that during a colectomy procedure, the scissor took more time than usual to cut.No particular mechanical resistance on the trigger.Then, after 10 minutes of use, the plastic attachment melted and got detached.The white plastic of the superior jaws was completely missing.The plastic attachment fell into the patient then was removed and kept.There was no patient consequence.
|
(b)(4).Batch # r94p39.Investigation summary: the analysis results found that the device was received with the tissue pad detached and returned in a plastic bag but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11, and the device did activate during functional testing.The device was disassembled to inspect the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|