It was reported that left hip revision surgery was performed.During the revision, the hemi head and modular sleeve removed.The bhr cup and short taper stem remained implanted.As of today, device return and additional information has been requested for this complaint but has not become available.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.The available medical documents were reviewed.Although the reported elevated cobalt & chromium, trunnionosis and pain are consistent with findings associated with metal debris, the root cause of the reported reactions cannot be confirmed.Further, it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|