• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122546
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Injury (2348); No Code Available (3191)
Event Date 07/13/2010
Event Type  Injury  
Event Description
It was reported that bilateral (this side left) hip revision surgery was performed due to unspecified reasons.
 
Event Description
It was reported that bilateral (this side left) hip revision surgery was performed due to pain, elevated test results and trunnionosis.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the hemi head and modular sleeve removed.The bhr cup and short taper stem remained implanted.As of today, device return and additional information has been requested for this complaint but has not become available.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.The available medical documents were reviewed.Although the reported elevated cobalt & chromium, trunnionosis and pain are consistent with findings associated with metal debris, the root cause of the reported reactions cannot be confirmed.Further, it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 46MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8685111
MDR Text Key147558864
Report Number3005975929-2019-00239
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2013
Device Catalogue Number74122546
Device Lot Number08KW19322
Date Manufacturer Received06/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ACETLR CUP HAP 52MM 74120152/093158; FEMORAL HEAD, PART AND LOT # UNKNOWN; MODULAR SLEEVE -4MM 12/14 74222100/08GW17828; SHORT TAPER STEM PC STD SZ 8 71932824/08FM09249
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age50 YR
-
-