Model Number N/A |
Device Problems
Flaked (1246); Peeled/Delaminated (1454)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 05/16/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a patient underwent an initial hip procedure.It was noted the surgeon broached to 9 stem but knew it was distally tight so he touched with a 10 broach.When he went to implant the stem, it sat very proud; he only tapped it 2 or 3 times.He pulled it out to broach it again and noticed the stem left coating in the femur.The surgeon thought the stem appeared over sprayed and that something was wrong.A new stem was opened and used without an issue.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was not confirmed.The returned products were visually and dimensionally inspected in accordance with documented procedures were found to be within specification.Review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|