The electrode was not returned to the service center for evaluation.As part of our investigation, an olympus sales representative visited the customer's site and observed 5-6 procedures using the facility's new generator and the phenomenon could not be reproduced.The sales representative also stated the generator that was used during the case has since been used but not with a resectoscope electrode.The facility was unsure about which device contributed to the doctor's incident.The cause of the reported event cannot be determined at this time.However, the a22257c ifu instructs users in the ¿inspection and testing¿ section ¿ inspect the instrument.Visually inspect the instrument.Warped electrodes, defective insulation and broken, cracked, or irregular cutting loops represent a danger for both the patient and the surgeon and must not be used.Never try to bend irregular cutting loops into shape!¿ if additional information is received, or the device is returned at a later this report will be supplemented accordingly.
|