The recipient reportedly experienced a sound quality issue.External equipment was exchanged, however, the issue did not resolve.The recipient ceased device use.On (b)(6) 2019, the recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The recipient is doing well.
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed near the fantail, and cut silicone overmold was observed on both top and bottom covers prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.The internal visual inspection noted silver migration across some electrical components.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|