Catalog Number 8065751763
Device Problem Complete Blockage (1094)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A surgeon reported the tubing line was clogged during a cataract procedure.The product was replaced with another one and there was no harm to the patient.
 
Manufacturer Narrative
Date received by manufacturer: 09/26/2019 04:20 pm.The lot complaint history was reviewed; this is the fourth complaint for the finish goods lot and fourth for this issue for this lot.The device history record shows the product was released per specifications.The returned sample was visually and functionally evaluated on a calibrated console.Inspection of the sample found no obvious defects that would have contributed to the reported event.The cassette was tested on a console, primed and tuned with the ultrasonic hand piece successfully and could achieve maximum vacuum.No system message was generated during functional testing.No fluid or air leaks, and no cracks were observed from the connectors.The irrigation and aspiration flow rates were measured and found to be within specifications.Fluid flowed from the balances salt solution (bss) bottle through the irrigation path continuously, no fluidic anomalies were observed.No occlusion or obstruction was identified during inspection and functional testing.The root cause of the customer's complaint could not be established as the returned fms cassette was evaluated and met specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8692657
MDR Text Key147948717
Report Number2028159-2019-01075
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2019
Date Manufacturer Received09/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK
-
-