It was reported that it became unable to measure cco with a hemosphere monitor on the first day of use of a swan ganz catheter.Although co had been measured with bolus mode, there was a difference observed between the cco value and the co value.The indicated value of both cco and co are unavailable.Also, neither rvp (right ventricular pressure) nor pap (pulmonary artery pressure) were unable to be measured although rap (right atrium pressure) could be measured at the time of the catheter insertion.The catheter and the cable were reconnected and then the problem was solved, but a dampened waveform of pressure was observed.No error message was displayed on the monitor.Following the above events, the monitor and the cable were checked and there was no problem observed.Patient demographic information requested but unavailable.There were no patient complications reported.
|
One catheter with attached monoject 1.5 cc limited volume syringe and four three-way stopcocks was returned for evaluation.A non-edwards contamination shield was located on the catheter body between the catheter tip and 70 cm proximal from the catheter tip.Returned product was evaluated using vigilance ii monitor.No fault messages appeared on the lab vigilance ii monitor when the catheter was connected.The thermistor was found to read 37.0 c when submerged into a 37.0 c water bath.The catheter ran cco in 37.0 c water bath on vigilance ii monitor for 5 minutes without error.The thermistor and thermal filament circuit were continuous, there were no open or intermittent conditions.No visible inconsistency was observed on eeprom data.Resistance value of the thermal filament circuit was within specification, measuring 38.67 ohms.Both the thermistor and thermal filament connectors were opened and no visible inconsistencies were found.No visible damage or inconsistency to the catheter body or balloon was observed.All through lumens were patent without any leakage or occlusion.The balloon inflated clear, concentric and remained inflated for more than 5 minutes without leakage.Balloon inflation test was performed using returned syringe with 1.5 cc air.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.Customer report of cco measurement issue and pressure measurement issue could not be confirmed during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.With any hemodynamic monitoring, patient parameters can change quickly and dramatically.Patient parameters should correlate with the patient¿s clinical manifestations.In this case it is unknown whether any user or procedural factors may have contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|