• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ TRUE SIZE THERMODILUTION CATHETER; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ TRUE SIZE THERMODILUTION CATHETER; SWAN-GANZ CATHETER Back to Search Results
Model Number 096F6J
Device Problems Inflation Problem (1310); Material Rupture (1546)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2019
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.0 cc limited volume syringe was returned for evaluation.Balloon was found to be ruptured at the central area and of the balloon latex.The ruptured edges were not able to match up.All through lumens were patent without any leakage or occlusion.No visible damage to the catheter body, balloon bonding sites or returned syringe was observed.Visual examination was performed under microscope at magnification 20x and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue was confirmed during the analysis.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user factors may have contributed to the event.This incident occurred before use so there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz true size thermodilution catheter did not inflate at the balloon leak test before use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ TRUE SIZE THERMODILUTION CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key8697887
MDR Text Key148425503
Report Number2015691-2019-02157
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
PMA/PMN Number
K810124
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/13/2020
Device Model Number096F6J
Device Lot Number61731998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2019
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-