• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770501
Device Problems Battery Problem (2885); Charging Problem (2892); Unexpected Shutdown (4019)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Supplemental will be submitted with evaluation results.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, the unit doesn't hold a charge, the battery shows one hundred percent and after ten minutes the scanner shuts off.
 
Event Description
Per biomed, the unit doesn't hold a charge, the battery shows one hundred percent and after ten minutes the scanner shuts off.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the unit doesn't hold a charge is confirmed.The scanner will shut down after the ac power is removed.The root cause of the reported issue is an internal li-ion battery failure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the unit doesn't hold a charge is confirmed.The scanner will shut down after the ac power is removed.The root cause of the reported issue is an internal li-ion battery failure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, the unit doesn't hold a charge, the battery shows one hundred percent and after ten minutes the scanner shuts off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8700326
MDR Text Key148179923
Report Number3006260740-2019-01667
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2019
Event Location Hospital
Initial Date Manufacturer Received 05/23/2019
Initial Date FDA Received06/14/2019
Supplement Dates Manufacturer Received07/17/2019
10/11/2019
Supplement Dates FDA Received08/09/2019
11/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2243072-10/11/2019-013-C
Patient Sequence Number1
-
-