It was reported that upon setting up for bilateral total knee, the surgical technician noticed 2 shims that were broken.The first shim size 5/6 10mm was missing the metal rings while the 5/6 12mm has several cracks starting at the center of the shim.This was noticed in time to get a new tray, surgery was not delayed.
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary: examination of the returned device confirms the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|