ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/24/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter stated that they received discrepant results for a total of 11 patient samples tested with the elecsys tsh assay, elecsys ft3 iii, and the elecsys ft4 iii assay on a cobas 8000 e 801 module.It was asked, but it is not known if any incorrect values were reported outside of the laboratory.This medwatch will apply to the ft4 assay.Please refer to the medwatch with patient identifier (b)(6) for information related to the tsh assay and refer to the medwatch with patient identifier (b)(6) for information related to the ft3 assay.The samples were initially tested on the reporter's e 801 analyzer on (b)(6) 2019.The samples were repeated on a centuar analyzer.The samples were also provided for investigation where they were tested using a cobas 6000 e 601 module and a cobas e 411 immunoassay analyzer on (b)(6) 2019 and a second e 801 analyzer on (b)(6) 2019.No adverse events were alleged to have occurred with the patient.The serial number of the e 801 analyzer used for investigation is (b)(4).Ft4 reagent lot number 380330, with an expiration date of 31-oct-2019 was used on this analyzer.The serial number of the e 601 analyzer used for investigation is (b)(4).Ft4 reagent lot number 391521, with an expiration date of 29-feb-2020 was used on this analyzer.The serial number of the e 411 analyzer used for investigation is (b)(4).Ft4 reagent lot number 378826, with an expiration date of 31-aug-2019 was used on this analyzer.
 
Manufacturer Narrative
The investigation determined the following for samples (b)(6): assays from different vendors can generate different values.This is related to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.The investigation determined the following for samples (b)(6): the investigation could not identify a product problem.The cause of the event could not be determined.The investigation determined the following for sample (b)(6): further investigations of the patient sample determined that it contained an interferent to a component of the ft3 assay.This limitation is covered in product labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8703330
MDR Text Key148171693
Report Number1823260-2019-02195
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/27/2019
Initial Date FDA Received06/17/2019
Supplement Dates Manufacturer Received05/27/2019
Supplement Dates FDA Received07/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-