The initial reporter stated that they have observed imprecision on quality controls and discrepant results for one patient sample tested with the elecsys ft4 iii assay lot 391521 on two cobas 8000 e 602 module analyzers.Since introducing the lot, the reporter states that they had to re-calibrate multiple times on both analyzers.No incorrect results were reported outside of the laboratory.The patient sample initially resulted with an ft4 value of 61.03 pmol/l when tested on the first e 602 analyzer.The sample was repeated on the second e 602 analyzer, resulting with a value of 54.19 pmol/l.The sample was also repeated on the first e 602 analyzer, resulting with a value of 49.21 pmol/l.No adverse events were alleged to have occurred with the patient.The serial number of the first e 602 analyzer is (b)(4) and the serial number of the second e 602 analyzer is (b)(4).
|