Analysis was approved for a generator replaced due to battery depletion, and the patient experiencing increased seizures.Upon receipt, the device was operating on low battery voltage and the elective replacement indicator (eri) flag was set.An open can measurement of the battery voltage confirmed that the eri flag had been properly set; confirming that the battery was ¿depleted¿.The device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Although the battery was depleted, monitoring results showed minor variations in the pulse generator¿s output signal and demonstrated that the device provided an expected level of output current for the monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.A comprehensive automated electrical evaluation showed that the pulse generator did not perform according to functional specifications.Both of the backup capacitor¿s to can measurements did not meet specification due to an open condition for the positive backup capacitors.Additionally, the positive backup capacitor anomaly affected the waveshape pulse overshoot measurement.After the positive backup capacitor was bridged device performed according to functional specifications.Visual analysis of the feed thru capacitors showed separation of the silver polyimide to the feed-thru wire connection causing "a considered open positive backup capacitor condition.¿ it is suspected that the positive backup capacitor was not electrically attached to the feedthru wire.Other than the noted open emi capacitor, there were no other performance or any other type of adverse conditions found with the pulse generator.No other relevant information has been received to date.
|