Model Number 8001000001 |
Device Problem
Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
Burn, Thermal (2530)
|
Event Date 02/28/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Customer did not make device available for evaluation.
|
|
Event Description
|
It was reported that a patient had burn marks on their legs that had the pattern of the blanket.The facility conducted an internal investigation and determined that the patient was burned due to liquid being left on the skin when the blanket was applied.The facility did not allege a malfunction of the device.The facility did not make the device available for evaluation by stryker.
|
|
Manufacturer Narrative
|
The facility informed stryker medical on (b)(6) 2019 that "there was no report of ulceration, any degree of burn, or blistering." they also noted that the patient was treated with a topical cream.Additionally, the facility did not notice any issue with the unit itself or any variations with flow and temperature.
|
|
Event Description
|
It was reported that a patient had burn marks on their legs that had the pattern of the blanket.The facility conducted an internal investigation and determined that the patient was burned due to liquid being left on the skin when the blanket was applied.The facility did not allege a malfunction of the device.The facility did not make the device available for evaluation by stryker.
|
|