Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 17june2019.A follow-up report will be submitted once the investigation has been complete.
 
Event Description
Customer reported bipap focus does not stay on.It is unknown if the device was in use at the time of the event or if there was a patient harm reported.
 
Manufacturer Narrative
Date received by manufacturer: 20 july 2019.Date of this report: 13 november 2019.The field service engineer (fse) replaced user interface board and internal battery.Ran battery charging test, reloaded system software, ran system calibration and ran performance verification.All testing passed.The determination could not be made that the device failed to meet specifications.It is unknown if the device was being used for treatment when the reported event occurred, and there is not relationship of the device to the reported problem.No parts were returned for failure investigation, therefore the root cause at the component level could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP FOCUS NONINVASIVE VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8706002
MDR Text Key148278610
Report Number2031642-2019-03893
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K053168
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/20/2019
Initial Date FDA Received06/17/2019
Supplement Dates Manufacturer Received05/20/2019
Supplement Dates FDA Received11/13/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse