Model Number TABLETOP
Device Problem Complete Blockage (1094)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported the instrument "blocked".There was no system message displayed.Event details and patient involvement are unknown.Additional information has been requested but not received.
 
Manufacturer Narrative
The company service representative examined the system but was unable to replicate the reported event.Unrelated to the reported event, the company service representative found system message (sm) [vitrectomy port transducers are out of tolerance.The vitrectomy probe is still available.Please contact field service] in the event log.The company service representative found that there was a significant crease with the footswitch cable causing poor contact.The pneumatic module and the footswitch cable were replaced to address those findings.Preventative maintenance (pm) was performed.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The system was found to have no problems; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8708360
MDR Text Key148644906
Report Number2028159-2019-01122
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/22/2019
Initial Date FDA Received06/18/2019
Supplement Dates Manufacturer Received07/25/2019
Supplement Dates FDA Received08/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-