• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2019
Event Type  malfunction  
Manufacturer Narrative
Patient weight was not available on the date of filing.Udi was not available on the date of filing.Pma/510(k) was not available on the date of filing.Manufacture date was not available on the date of filing.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system.It was reported that intra/peri-operatively during the select patient/acquire exams task of a fess procedure, the clinical specialist (cs) loaded an exam and it was inverted.Technical services (ts) provided the pixel offset settings, it was not known if that resolved the issue at the time of the event.There was a delay to the procedure of less than one hour.There was no reported impact on patient outcome.
 
Manufacturer Narrative
Patient weight was received.Updated with resolution.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received: adjusting the pixel offset settings resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8708606
MDR Text Key148322072
Report Number1723170-2019-03528
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/13/2019
Initial Date FDA Received06/18/2019
Supplement Dates Manufacturer Received06/18/2019
Supplement Dates FDA Received07/09/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age70 YR
Patient Weight59
-
-