• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, BRASIL LTDA. EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, BRASIL LTDA. EPIC SUPRA VALVE W/FLEXFIT; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number ESP100-23
Device Problem Material Puncture/Hole (1504)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2019, a 23mm epic supra valve w/flexfit was selected for impaling and during device preparation a small hole was detected in the valve and caused extended procedure time.The valve was not used and no patient consequences were noted.
 
Manufacturer Narrative
The reported event that there was a small hole in the valve could not be confirmed.No anomalies were found with the valve cusps, including no holes, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper cuspal coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC SUPRA VALVE W/FLEXFIT
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
MDR Report Key8709741
MDR Text Key148413499
Report Number3001883144-2019-00052
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/12/2023
Device Model NumberESP100-23
Device Catalogue NumberESP100-23
Device Lot Number6927366
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2019
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-