MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74121146
Device Problems Material Fragmentation (1261); Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273)
Patient Problems Failure of Implant (1924); Pain (1994); Injury (2348); No Code Available (3191)
Event Date 03/12/2018
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed due to clicking, pain and the presence of pseudotumor as a result of metallosis.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the bhr head was removed.The bhr cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and bhr head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The noted intraoperative findings of pseudotumor may be consistent with findings associated with metallosis; however, without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported pseudotumor cannot be confirmed, and it cannot be concluded that the reported reaction was associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be re-opened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 46MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8712209
MDR Text Key148432196
Report Number3005975929-2019-00251
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2012
Device Catalogue Number74121146
Device Lot Number73754
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/17/2019
Initial Date FDA Received06/19/2019
Supplement Dates Manufacturer Received04/23/2020
Supplement Dates FDA Received04/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74120154 ACETLR CUP HAP 54MM W/ IMPTR 72465; UNKNOWN BHR DEVICE/UNKNOWN LOT; UNKNOWN BHR DEVICE/UNKNOWN LOT
Patient Outcome(s) Hospitalization; Required Intervention;
-
-