• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40C
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Therapeutic Effects, Unexpected (2099); Palpitations (2467)
Event Date 04/06/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient received inappropriate shock from the implantable cardioverter defibrillator.Just prior to the shock, patient felt lightheaded with retrosternal chest pressure.The shock put her back into sinus rhythm.The cardiologist stated that though the abnormal rhythm prior to her shock was electronically interpreted as ventricular tachycardia, it really was at with some atrioventricular nodal reentry tachycardia and atrial fibrillation.No intervention was performed for inappropriate shock.The patient had no further issues from the shock.The patient went to the ed on (b)(6) 2019 for other issues, and the device check done on (b)(6) 2019 showed that no further episodes of atrial arrhythmias since the (b)(6) 2019 incident.The patient was stable.
 
Event Description
New information states that the patient experienced palpitations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8713910
MDR Text Key148475926
Report Number2938836-2019-04901
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508261
UDI-Public05414734508261
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Model NumberCD3365-40C
Device Catalogue NumberCD3365-40C
Device Lot NumberA000007769
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/07/2019
Initial Date FDA Received06/19/2019
Supplement Dates Manufacturer Received08/20/2019
Supplement Dates FDA Received09/05/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient Weight78
-
-