• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0302
Device Problems Difficult to Insert (1316); Deformation Due to Compressive Stress (2889)
Patient Problem Perforation (2001)
Event Date 05/30/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed for perforation of the right iliac vein requiring percutaneous repair.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.During insertion of the steerable guide catheter (sgc), resistance was noted while advancing the catheter through the femoral vein.It was observed on fluoroscopy that the loop of the guidewire was causing the sgc to be kinked in the right iliac vein.Upon sgc removal, damage to the right iliac vein was observed on fluoroscopy; a leak was seen.The patient was stable and the lesion was successfully repaired percutaneously with covered stents.The mitraclip procedure was aborted.Post procedure mr remained at grade 4.On (b)(6) 2019, a mitraclip was successfully implanted reducing mr from 4 to 1.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history no similar incidents reported from this lot.The reported difficulty inserting the sgc appears to be related to the kink in the sgc and the kink was due to the loop of the guidewire (procedural conditions).The reported patient effect of perforation, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The perforation was due to procedural conditions.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key8714553
MDR Text Key148526364
Report Number2024168-2019-04885
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/10/2020
Device Catalogue NumberSGC0302
Device Lot Number90311U131
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/30/2019
Initial Date FDA Received06/19/2019
Supplement Dates Manufacturer Received08/22/2019
Supplement Dates FDA Received08/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-