BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G125 |
Device Problems
Failure to Capture (1081); High impedance (1291); Loose or Intermittent Connection (1371); Over-Sensing (1438); Pacing Problem (1439); Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/23/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that during a left ventricular (lv) lead replacement procedure there was no capture, oversensing, pacing inhibition and high out of range impedance seen on this cardiac resynchronization therapy defibrillator (crt-d) system.Furthermore, there were telemetry issues so the crt-d was removed.There were also concerns of possible header damage, loose connection and setscrew issues.The patient was successfully placed with a new lv lead and crt-d and no further issues were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Event Description
|
It was reported that during a left ventricular (lv) lead replacement procedure there was no capture, oversensing, pacing inhibition and high out of range impedance seen on this cardiac resynchronization therapy defibrillator (crt-d) system.Furthermore, there were telemetry issues so the crt-d was removed.There were also concerns of possible header damage, loose connection and setscrew issues.The patient was successfully placed with a new lv lead and crt-d and no further issues were reported.
|
|
Search Alerts/Recalls
|
|
|