• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; BAG, URINARY, ILEOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; BAG, URINARY, ILEOSTOMY Back to Search Results
Model Number 650832
Device Problem Malposition of Device (2616)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event Description
It was reported the starter hole was off center in 16 wafers which is causing leakage between the mass and bag.No photo provided.
 
Manufacturer Narrative
Batch record review: lot 8d00904 was manufactured on 04/09/2018, in the convex 1pc line with a total of (b)(4) ea.A batch record review was performed to verify if all the applicable procedures were followed and no issues were found, all the components for assembly were correct per bom and all the tooling information documented was also correct, under icc code 650832, sap material id 1004090 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.The crew requirements and responsibilities, process parameters, quality and in-process inspections, line operations, process troubleshooting and relevant documents to the process were run according to process instructions pi21-076.Photograph, video and/or physical sample evaluation: no photograph associated with this case were received.No unused return sample has been received for this complaint.Conclusion summary of the related event: based in the investigation findings, the root cause identified for the issue ¿wafer off center in the pouch¿, reported under failure mode ¿ost-pmc1.8 skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur¿ is attributed to: 1) machine: the investigative process concluded that all the machinery/ tooling items used in the manufacturing process complied when compared against drawing and pi21-076 rev 37.0 specifications; however, as part of this investigation on 10/may/2019, maintenance and facilities manager (aurys arias) and senior process engineer (alejandro sosa) performed a deep assessment to the sub-assembly station of the convex 1pc machine identifying the conditions causing the wafer off center and the possible solutions to correct the failure.As conclusion of the assessment, the chain that coordinates flange loading station with adhesive loading station needed lubrication and the chain and other components of the mechanism needed a corrective maintenance.In order to maintain this condition in the long term, a deeper intervention of the mechanism is required.Furthermore, regarding the rotary table station of the convex 1pc machine, it was identified that due to the demands of the process, the tooling (cup pouch uro 177a) used to manufacture the urodress product family require a redesign modification to reduce the variability within the manufacturing process.As a contributor factors the following opportunities for improvements were identified: 2) process / method investigation: a) reference in the pi21-076; section 8.0 ¿quality instructions¿ the applicable qc tooling used for product¿s quality inspection purpose of pouch uro 177a products family in order to address manufacturing personnel the use of them.B) identification of the defect during the manufacturing process, it should be considered the implementation of a job aid for product centralization out of specification.3) process / method / measurement investigation a) introduce qc tooling used for product¿s quality inspection purpose in the calibration program to guarantee measurement assurance.4) manpower: a) there are not manpower causes associated with the root cause; even though, opportunity to improve the loading of the adhesive disc during the manufacturing process was identified by implementing a standard work instruction for convex 1pc to indicate manufacturing operators how to properly load the adhesive disc.No issues were identified for material and environment investigations.A capa plan was generated for the mitigate the root causes identified.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Type of Device
BAG, URINARY, ILEOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key8730988
MDR Text Key149082823
Report Number9618003-2019-04492
Device Sequence Number1
Product Code EXH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/18/2023
Device Model Number650832
Device Lot Number8D00904
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/17/2023
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient SexMale
-
-